Standard Content
LUFT3
1 equipment
1 articulated arm
1 trolley
1 air filter
1 air hose
1 O2 hose
1 patient circuit
1 test lung
1 touchscreen monitor
2 exhalation valves (1 extra and 1 with the equipment)
1 user manual
1 technical manual
1 proximal flow sensor
1 oxymetry sensor
1 capnography sensor
Tech Specs part1
VENTILATION MODES
PATIENT RANGE: Adult and Pediatric
- VCV - Volume Controlled Ventilation
- PCV - Pressure Controlled Ventilation
- PRVC - Pressure Regulated Volume Control
- PSV - Pressure Support Ventilation
- CPAP - Continuous Positive Airway Pressure
- NIV - Non-Invasive Ventilation
- SIMV (VCV / PCV) + PSV - Synchronized Intermitent Mandatory Ventilation with Pressure Support
- MMV + PSV - Mandatory Minute Ventilation with Pressure Support
- PSV + Tv - Pressure Support Ventilation with Assured Tidal Volume
- Biphasic Pressure
- PCV - Pressure Controlled Ventilation
- Continuous Flow
- PSV - Pressure Support Ventilation
- CPAP - Continuous Positive Airway Pressure
- Nasal CPAP
- SIMV (PCV) + PSV - Synchronized Intermitent Mandatory Ventilation with Pressure Support
FEATURES
- Backup Ventilation:
PCV or VCV in adult, pediatric and neonatal
Emergency Ventilation - FiO2: 21 to 100% (electronically regulated on the panel with graphic display monitoring)
- Inspiration Time: 0.10 to 30.0 seconds
- I:E Ratio: 5:1 - 1:99
- Frequency: 1 - 180 cpm (cycles per minute)
- Tidal Volume: 5 - 2500 ml
- Sensibility: 0.2 - 15 l/min (by flow); -0.5 to -15 cmH2O (compensated PEEP)
- PCV: 1 to 80 cmH2O over PEEP (with adjustable “Rise Time”)
- PSV: 0 to 80 cmH2O over PEEP (with adjustable “Rise Time”)
- Inspiration Pressure: 2 to 120 cmH2O
- Rise Time: 5 levels
- Expiration Sensibility: Adjustable from 5 to 80% of the initial flow
- PEEP/CPAP: 0 - 50 cmH2O
- Nebulization: 1 to 20 min. with automatic inspiratory volume compensation
- TGI: Synchronized with the expiration phase
- Inspiratory Flow:
In VCV mode: Automatic Adjustment.
In PCV and PSV modes: Up to 180 l/min.
Continuous flow in Neonatology: 2 to 15 l/min.
Inspiration flow in Neonatology: 2 to 30 l/min. base Flow: Off up to 50 l/min.
- Expiratory Flow: Up to 120 l/min.
- Sigh: Cycles per hour, quantity, maximum tidal volume and manual trigger.
- Inspiratory Pause: 0.2 - 2 seconds
- 100% FiO2: Oxygenation for aspiration with synchronized system – 1 to 20 min.
- Flow waveform: Square; 100% Descending ramp; 50% Descending ramp; Sinusoidal; Ascending ramp
- Automatic Air - O2 gas net bypass: In case of failure in one of them, the equipment continues operating normally.
- Inspiratory Pressure Inner Safety Valve: Adjusted in 120 cmH2O
- Regulatory Pressure Valve of Air and O2 input: internally incorporated in the equipment
- Stand by: To hold the set configuration without cycling
- Scales: Automatic updated for vertical and horizontal readings
- Freeze: To hold graphics on the screen while the ventilator keeps working normally
- RS232 signal connector: For external communication to software and signals input
- USB signal connector: For technical service and software update
OUTPUT PARAMETERS MONITORING
- Airway Pressure: Peak, Plateau, Mean, base (PEEP)
- Inspiratory time
- Exhalation time
- I:E ratio and spontaneous breathings
- Inspiratory pause
- Inspired/exhaled tidal volume (distal/proximal)
- Inspiratory/exhalation peak flow (distal/proximal)
- Dynamic compliance
- Total frequency
- Exhalation time constant
- Graphic indicator of spontaneous and mechanical cycles
- Minute volume (distal/proximal)
- FiO2 concentration
- Compressible volume
- Ti/Ttot relation
- etCO2
- Inspired CO2
- Total leakage
- AutoPEEP
- Dynamic/Static compliance
- Static inspiratory/expiratory compliance
- Slow vital capacity
- P0.1 (airway obstruction pressure)
- P-V curve with low flow
- Tobin index
- Stress index
ALARMS
- Inspiratory max/min airway pressure (disconnection of the patient circuit)
- Max/min exhaled tidal volume
- Max/min exhaled minute volume
- Apnea with adjustable time
- Max respiratory frequency
- Max/min PEEP and continous pressure
- Gas supply source (air-oxygen)
- Power outage
- Low battery
- Microprocessor
- Interrupted cycle in pressure modes
- Max/min FIO2
- Inverted I:E ratio
GRAPHICS
- Pressure - time
- Flow - time
- Volume - time
- CO2 - time
- Volume loop - pressure
- Flow loop - volume
- Pressure loop - flow
- CO2 loop - volume
UP TO 24h TENDENCIES
- Peak and base pressure
- Flow
- Tidal volume
- Minute volume
- Frequency
- Dynamic compliance
ALARMS LOG
- Records the last 1000 events
OTHER MENUS
- Menu for indication of hours of use and performed services
- Altitude adjustment
- Language selection
- Audio volume adjustment
Tech Specs Part2
SELF-TESTS
- Altitude compensation
- Capture of atmospheric pressure
- Oxygen cell detection
- Verification of hours of use
- Sensors calibration
- Proximal sensor detection and calibration
- Circuit leakage measuring
- Circuit compliance measuring
- Calibration of the exhalation flow sensor
- Calibration of the oxygen cell
- Calibration of the internal flow sensors
- Air and oxygen proportional valves tests
- PEEP control valve tests
- Detection and initialization of the capnograph
OTHER SAFETY CHARACTERISTICS
- Automatic compensation of the gases
- Warning of maintenance need
- Possibility of working without flow sensor
- Possibility of working without proximal sensor
- Possibility of working without oxygen cell