Specification | Ventilator Microprocessor controlled system, suitable for adult & pediatric, LED display, convenient using, humidifier use electronic knob setting. Main Unit : Patient type: Adult & pediatric Driven mode: gas driven, electronically controlled Working mode: pressure limited, time cycled, volume controlled Setting: touch key Ventilation mode: IPPV, SIPPV, SIGH, SIMV, MAN, SPONT Tidal volume: 0~1200ml Breathing rate: 6~60bpm I:E ratio: 2:1~1:4 Trigger Sensitivity: -2Kp~2Kpa ASB pressure: 0~6Kpa High pressure:1~8Kpa Low pressure: 0~2Kpa High MV: 0~30L/min Low MV: 0~15L/min Silence time for alarm: 120s Monitoring: tidal volume, minute volume, airway pressure Alarm: high/low airway, high/low MV Power supply: 220V 10%, 50Hz 1Hz Humidifier : Temperature control High temperature alarm, auto-power cut off Power supply: 220V 10%, 50Hz 1Hz
|